Model Number N/A
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Pain (1994)
Event Date 10/09/2023
Event Type  Injury  
Event Description
It was reported that the patient underwent a right knee arthroplasty to address tibial loosening and pain that increased with weight-bearing approximately four (4) years post-operatively.Attempts have been made, however, no additional information is available.
 
Manufacturer Narrative
(b)(4).D10 - concomitant devices - persona partial knee articular surface right medial size g 8mm catalog #: 42528200708 lot #: 64157607, persona partial knee femoral component cemented size 3 right medial catalog #: 42558000302 lot #: 63982592, biomet bone cement r 1x40 us catalog #: 110035368 lot #: 824fac2202.The complainant has indicated that the product will not be returned to zimmer biomet for investigation, as the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this patient; please see all reports associated with this event: 0001825034-2023-02598.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Radiographic review identified possible polyethylene wear, however, no tibial loosening was noted.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this patient; please see all reports associated with this event: 0001825034-2023-02597, 0001825034-2023-02598, 0001825034-2024-00651.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA PARTIAL KNEE TIBIAL COMPONENT CEMENTED SIZE G RIGHT MEDIAL
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18074920
MDR Text Key327423987
Report Number0001825034-2023-02597
Device Sequence Number1
Product Code HSX
UDI-Device Identifier00880304812734
UDI-Public(01)00880304812734(17)270630(10)63720353
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42538000702
Device Lot Number63720353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
Patient Weight96 KG
-
-