Device evaluation: one device was received.A visual inspection found that the microswitch was alarming, the pump was not working, the enclosure was cracked under quick connect and quick connect was corroded.During the functional testing, it was found that the microswitch was alarming with the disposable.The cause of the issue was found to be a faulty microswitch.The microswitch was replaced along with the pump, enclosure, quick connect, reservoir gasket and o-rings.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.For all enquiries or follow-up questions related to the record, do not use (b)(6) located in sections g.1., please direct those to the following: (b)(6).
|