The dispatched fse checked the device on-site and could confirm the reported aspect of ventilator failure.The blower unit was replaced; the device passed all consecutive tests and was returned to use.The log file was evaluated by the manufacturer - the entries indicate that the device suddenly lost the signal of one of the hall sensors of the blower unit.These sensors are necessary to control the motor run.As designed the workstation performed a shut-down of automatic ventilation to prevent from potentially hazardous out put and/or serious damages and alerted the user to this condition by means of a corresponding alarm.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|