Catalog Number MK06000
Device Problems Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2023
Event Type  malfunction  
Manufacturer Narrative
The dispatched fse checked the device on-site and could confirm the reported aspect of ventilator failure.The blower unit was replaced; the device passed all consecutive tests and was returned to use.The log file was evaluated by the manufacturer - the entries indicate that the device suddenly lost the signal of one of the hall sensors of the blower unit.These sensors are necessary to control the motor run.As designed the workstation performed a shut-down of automatic ventilation to prevent from potentially hazardous out put and/or serious damages and alerted the user to this condition by means of a corresponding alarm.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that the device posted a ventilator failure during use.The device was replaced; no patient consequences were reportedly resulting from the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSEUS A500
Type of Device
ANESTHESIA UNIT
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key18090475
MDR Text Key327603194
Report Number9611500-2023-00398
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675253600
UDI-Public(01)04048675253600(11)230517(93)MK06000-42
Combination Product (y/n)N
PMA/PMN Number
K133886
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMK06000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/31/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-