Device Problem Fire (1245)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2023
Event Type  malfunction  
Event Description
It was reported that there was an odor and small amount of smoke in the respiratory icu.The ventilator in patient use was found to have a burned capacitor of the air pump.There was no patient injury.
 
Manufacturer Narrative
Ge healthcare âs investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: patient information could not be obtained due to country privacy laws.Block e1: the initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Block d4 unique identifier:(b)(4).Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718 h3 other text : device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
The end user confirmed it was the capacitors from the air compressor that generated smell and smoke.After engineer replaced the non-ge capacitor, there was no further issue with the device.H3 other text : the end user confirmed it was the capacitors from the air compressor that generated smell and smoke.After engineer replaced the non-ge capacitor, there was no further issue with the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESCAPE R860
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key18093439
MDR Text Key327648781
Report Number2112667-2023-05864
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K210384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/26/2024
Date Device Manufactured11/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-