Resuscitation of the patient was initiated after cardiopulmonary arrest and the spur ii was found not to be inflating due to a tear.A backup spur ii was retrieved to continue the resuscitation.Follow-up with the customer revealed that the patient expired.During the manufacturing process all the resuscitators undergo a function test, consisting of an inspiratory/expiratory test, reservoir bag test and high/low pressure test, in order to verify that the finished products meet the design specifications.A tear on a product would have been detected during these tests, which take place prior to packing.Since there was no sample nor picture available for investigation, the reported failure could not be verified and we can only speculate on the position of the tear.The tear could have been on a compressible bag or reservoir bag and in both cases such tear could be a result of reprocessing or cutting through the wall by a sharp object.Additionally, by pulling onto the reservoir bag or the bag not welding well could result in the tear as well.We can only exclude reprocessing and reuse from the listed potential causes based on the limited information, thus the root cause could not be determined.The accompanying instructions for use state 'always inspect the resuscitators and perform functional test after unpacking, cleaning, assembly and prior to use'.Tear in the compressible or reservoir bags is detectable during pre-check.This is the second complaint within the past 12 months.The root cause is unknown, however we will keep monitoring the occurrence of similar incidents.
|