• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ACCU-CHEK INFORM II TEST STRIPS; BLOOD GLUCOSE MONITORING TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ACCU-CHEK INFORM II TEST STRIPS; BLOOD GLUCOSE MONITORING TEST STRIPS Back to Search Results
Catalog Number 05942861001
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2023
Event Type  malfunction  
Manufacturer Narrative
The accu-chek inform ii meter serial number was (b)(6).The reporter's meter and test strips were requested for investigation.The products have not been received at this time.If the products are returned in the future, a follow-up report will be submitted.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Event Description
The reporter complained of discrepant glucose results for 1 patient's sample tested on accu-chek inform ii meter compared to an unknown laboratory method.At 6:51 p.M.The meter result was 510 mg/dl.The medical staff did not believe the result and they took a venous sample from the patient and did a laboratory test.The laboratory result was 72 mg/dl.The time difference between the test measurements was within 15 minutes.Qc was performed on the day of the event and it was acceptable.
 
Manufacturer Narrative
The component code was updated.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18100981
MDR Text Key327728059
Report Number1823260-2023-03564
Device Sequence Number1
Product Code LFR
UDI-Device Identifier00365702428102
UDI-Public00365702428102
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05942861001
Device Lot Number670850
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
Patient Weight89 KG
-
-