Model Number N/A |
Device Problems
Corroded (1131); Unstable (1667); Calibration Problem (2890); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the motor speed was unstable, calibration issue, worn reciprocation arm and corroded motor.The event timing was during preventative maintenance.There is no harm or delay reported.No adverse events were reported as a result of this malfunction.Due diligence is complete and there is no additional information available.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.G2: foreign country - france.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
Review of the most recent repair record determined that the unit was out of calibration, arm was worn and that the motor failed.The motor and reciprocation arm were replaced, the device was recalibrated, and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|