• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Calibration Problem (2890); Material Twisted/Bent (2981); Naturally Worn (2988); Patient Device Interaction Problem (4001)
Patient Problem Laceration(s) (1946)
Event Type  Injury  
Event Description
It was reported that during surgery the skin graft did not harvest properly using a three-inch guard resulting in a poor-quality skin graft.The taken graft was damaged as the central piece of skin was missing in the graft, leaving only two small strips.An additional graft was taken which led to a larger donor site wound on the patient's thigh.There was no delay noted in the procedure.It was also reported that all of the hospitals' dermatome kits are old and none of them have been serviced/maintained.Due diligence is complete and there is no additional information available.
 
Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.E1 : (b)(6).G2 country: united kingdom.Multiple mdr reports were filed for this event, please see associated reports: 0001526350-2023-01170-1.
 
Manufacturer Narrative
This mdr is a duplicate of 0001526350-2023-01153.The initial report was created in error and should be voided.
 
Event Description
This mdr is a duplicate of 0001526350-2023-01153.The initial report was created in error and should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18102160
MDR Text Key327737682
Report Number0001526350-2023-01485
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexPrefer Not To Disclose
-
-