Model Number N/A |
Device Problems
Unstable (1667); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial report.G2: foreign: france.Customer has indicated that the product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during preventative maintenance the device needed to be repaired for a motor issue, worn reciprocation arm, and unstable motor speed.There was no harm or injury reported.Due diligence is complete.As no additional information is available, we are unable to provide further information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the reciprocation arm was worn and the motor speed was unstable.The motor and reciprocation arm were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information is available.
|
|