Model Number V60
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Event Description
Philips received a complaint from the customer on the v60 indicating that the standby button was not active, and that the device will not go into standby mode.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The rse performed a troubleshoot with the customer and confirmed the device was showing 15 for the air flow slpm when it should be -1 or 1.The rse then advised the customer that a replacement of the flow sensor assembly is recommended to resolve the issue.The rse provided the customer with the part number for the flow sensor assembly to order and replace.
 
Manufacturer Narrative
Multiple attempts have been made to try to obtain further information about this case but no response has been received from the customer.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18104934
MDR Text Key328500860
Report Number2518422-2023-29732
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020047
UDI-Public00884838020047
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053616
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/11/2023
Date Device Manufactured07/25/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-