The initial reporter stated they received questionable results for one patient sample tested with elecsys pth on two cobas e 801 module analyzers.Refer to the attachment for all patient data.The sample was initially tested on the customer's e 801 analyzer on (b)(6) 2023.The customer repeated the sample with x2, x4, and x8 dilutions.The patient's previous pth values ranged between 50 and 100 pg/ml, so the results from the complained sample were questioned.The sample was provided for investigation where it was tested with pth and pth (1-84) on a second e 801 analyzer on (b)(6) 2023.Pth and pth (1-84) were also re-measured after x2, x4, and x8 manual dilutions with both low pth patient serum and diluent solution.
|