A facility reported that during "craniotomy for aneurysm clipping" procedure, the surgeon positioned the patient in the mayfield modified skull clamp (a1059) and adjusted the torque screw to 60 lbs.While performing the surgery, the surgeon heard a noise that raised concerns about the swivel arm potentially unlocking, although no visible movement was observed.The surgery continued as planned, and after completing the aneurysm clipping, they broke scrub and surgeon evaluated the mayfield clamp, which was still in use.The clamp's pressure had decreased to 20 lbs., but the rocker arm remained locked.Fortunately, this incident resulted in only a 10-minute delay in the surgery, which had no adverse effects on the case, and the patient did not suffer any injuries.
|
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit exhibited some slight movement in the lock which needed to be adjusted to remove any movement.However, the reported issue could not be duplicated.When the unit is properly positioned and put under pressure, it would not lose pressure.All worn components were replaced with new parts, and general cleaning and maintenance were performed.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.Probable root cause is improper or suboptimal positioning of the patient.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|