EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number D97120F5
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.The reported pacing issue was confirmed.The continuity testing found that a short condition occurred between the proximal and distal circuits in the y adaptor.There was no open or short condition observed in the leadwires between distal side of y adaptor and the electrodes.There was no visible damage observed from catheter body, balloon, and windings.The balloon inflated clear and concentric and remained inflated for five minutes without leakage.A device history record review was completed and documented that device met all specifications upon distribution.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that during use after catheter insertion, it was unable to pace.The issue was resolved by replacing the catheter.The following information was unknown regarding the background of malfunction occurrence, what kind of surgery or examination the catheter was used for, or whether the patient had cardiac conduction defect.The patient demographic information was requested but not available.There were not patient complication reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
9492017706
MDR Report Key18108048
MDR Text Key328630592
Report Number2015691-2023-17390
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K822723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD97120F5
Device Lot Number64752350
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-