Model Number D97120F5 |
Device Problem
Pacing Problem (1439)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.The reported pacing issue was confirmed.The continuity testing found that a short condition occurred between the proximal and distal circuits in the y adaptor.There was no open or short condition observed in the leadwires between distal side of y adaptor and the electrodes.There was no visible damage observed from catheter body, balloon, and windings.The balloon inflated clear and concentric and remained inflated for five minutes without leakage.A device history record review was completed and documented that device met all specifications upon distribution.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
|
Event Description
|
It was reported that during use after catheter insertion, it was unable to pace.The issue was resolved by replacing the catheter.The following information was unknown regarding the background of malfunction occurrence, what kind of surgery or examination the catheter was used for, or whether the patient had cardiac conduction defect.The patient demographic information was requested but not available.There were not patient complication reported.
|
|