• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number ROTAFLOW CHINESE ZH-PLUG
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.Further information has been requested but has not yet been received.A follow up medwatch will be submitted when additional information becomes available.
 
Event Description
The event occurred in china during patient treatment.It was reported that after being disconnected from ac power, the rotaflow would unintentionally shut down while being used in battery mode.The rotaflow was reconnected to ac power and treatment continued as intended, with no consequences to the patient.No harm to any person has been reported.Complaint id (b)(4).
 
Event Description
Complaint id (b)(4).
 
Manufacturer Narrative
The event occurred in china during patient treatment.It was reported that after being disconnected from ac power, the rotaflow would unintentionally shut down while being used in battery mode.The rotaflow was reconnected to ac power and treatment continued as intended, with no consequences to the patient.No harm to any person has been reported.A getinge service technician investigated the rotaflow console with serial number (b)(6).The technician confirmed the reported failure and determined that the "battery pack with fuse" (material# 70101.7188) will need to be replaced.A similar complaint has previously been investigated by life cycle engineering (lce), and the most probable root cause of the error is that the battery had an increased internal resistance.This could be caused by incorrect maintenance habits.But the actual cause in this case cannot be determined.Based on these investigation results the reported failure "battery failure" could be confirmed.A device history record (dhr) review was performed on 2023-11-22.There is no indication that manufacturing issues occurred, thus production related influences are unlikely to have contributed to the reported failure.In order to avoid reoccurrence of the reported failure, the customer will be informed by the getinge sales and service unit (ssu) to follow the chapter in the instruction for use heart-lung support system rotaflow system| 4.4 | en | v15.Chapter 3.3.4.Check battery capacity every 6 months, at the latest.The battery must be replaced by the authorized technical service every 2 years, at the latest.The battery must be replaced sooner if it cannot be fully charged within 8.5 hours or if the system cannot be operated with the fully charged battery.Chapter 5.6.1.Before starting the application, check the points listed in "check before every use".Before each use, ensure that the batteries are fully charged.If the battery capacity is low an acoustic signal sounds on the device.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18110153
MDR Text Key329202601
Report Number8010762-2023-00562
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW CHINESE ZH-PLUG
Device Catalogue Number701046405
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-