SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT, EXPANDABLE; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C45101344-NLJ |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: b3: month and year of event have been provided, day is unknown.D4: udi section is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that during pre-testing, there was a crack in the circuit connection.No adverse effects have been reported.
|
|
Manufacturer Narrative
|
One device and three photos were returned for investigation.Per visual inspection of the device, a broken connector was found.The reported complaint was confirmed, and this problem causes an air leak in the circuit.No corrective action was taken.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
|