ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 146870489 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is not available for investigation as the customer has discarded it.Without the return of the device, a probable cause is unable to be determined.If the device becomes available a supplemental report will be submitted.E1 initial reporter phone number: (b)(6).
|
|
Event Description
|
The event occurred on an unspecified date and involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inc where it was reported that the tubing breaks easily.The customer reported to see if the adhesive coming apart was easily replicable.They were not able to easily replicate with the same product.The customer was easily able to replicate with minimal force with this lot number that they had in stock at the time of the event.During their attempt to replicate, they noticed that the adhesive responds differently in different products.They were able to replicate with most sets available on site but not of the lot that the event occurred with.The product has been used before it was noticed.No patient harm was reported.
|
|
Search Alerts/Recalls
|
|
|