This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: e3: reporter is a (b)(6) sales representative.Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: functional : heat cable, electrical.Per service reports, this complaint can be confirmed.The device was returned to customer unrepaired as system was not performing to specification.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.The faulty parts was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
It was reported from the sales rep in colombia that during a knee arthroscopy surgical procedure on 09/15/2023 the micro tornado hp w handcontrol device did not react when the operating buttons are pressed.It was reported that the blade was opened but could not be used because the handpiece did not work therefore it could not be charged.During an in-house engineering evaluation, it was determined that the device had the following defect: functional: heat cable, electrical.There was patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|