Correction: d9/h3, h6 method, results, conclusion the reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.The device inspection revealed the following: visual inspection: a visual inspection of the device shows that the foot and main body are in good condition and that hammer pad 1 and hammer pad 2 have a normal amount of impact deformations from use.Functional inspection: a functional inspection of the device showed that it was functioning normally.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause cannot be established as the device is fully functional and conforms to specification.If any further information is provided, the complaint report will be updated.
|