AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1TH190
Device Problems Corroded (1131); No Display/Image (1183); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2023
Event Type  malfunction  
Event Description
The customer reported to olympus that the evis exera iii bronchovideoscope no image, monitor had shown only black color which happened before the case during preparation for use.The case was completed using a similar device.Upon inspection and evaluation, due to corrosion on suction connector, the image is not displayed.There was no report of patient harm or user injury associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunctions found during incoming inspection and evaluation.
 
Manufacturer Narrative
The device was returned to olympus for inspection, and the customer's reported issue mentioned in b5 was confirmed.The following findings were also noted during device evaluation: connecting tube has a scratch.Connecting tube has discoloration, due to damage on channel-tube, water tightness is lost, and bending section cover has corrosion.Based on the results of the investigation, olympus confirmed the biopsy channel was leaking while the user was handling the device, and water invaded the device.Due to the water invasion, abnormality of electronic parts occurred which led to the malfunction.A device history review revealed no issues that could have caused or contributed to the reported issue.This issue is addressed in the instructions for use (ifu): warnings and cautions: caution.Turn the video system center on only when the endoscope connector is connected to the video system center.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.Warnings and cautions: disappeared or frozen endoscopic image: warning.Follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system.Inspection of the endoscopic image.Confirm that the wli and nbi endoscopic images are normal.5.1 troubleshooting: if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18117427
MDR Text Key329318091
Report Number9610595-2023-17007
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K121959
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-1TH190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-