|
Model Number A22040A |
Device Problem
Scratched Material (3020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the black lock part came off the inner sheath, for 26 fr.Outer sheath.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: scratches on the inside of the tip beak.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found the black lock part came off, there were scratches on the inside of the tip beak, the seal rubber was discolored and there was tensioning corrosion.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 16 years since the subject device was manufactured.Based on the results of the investigation, root cause of the reported damage is consistent with wear from use over time and improper handling.However, specific root cause could not be determined.The issue can be detected/prevented by following the instructions provided in: the instructions for use, chapter 4 before use ¿ warning infection control risk - properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing - inspecting the product - visually inspect the product.Make sure that it has: no corrosion no dents no scratches ceramic insulation at distal end - visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning - risk of injury - impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product - if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|