Model Number 71992-01 |
Device Problem
High Readings (2459)
|
Patient Problems
Hypoglycemia (1912); Shaking/Tremors (2515); Increased Appetite (4570)
|
Event Date 11/03/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high readings issue was reported with use of the abbott diabetes care (adc) device.Customer received unspecified higher sensor scan results when compared to unspecified readings obtained on built-in meter.As a result, customer experienced "feeling of hunger", "shaking", and received third-party treatment of "juice and sugar" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.Built-in result(s) of 104 mg/dl and [comparison type] result(s) of 428 mg/dl were plotted on a parkes error grid and fell into the "d" zone, showing the difference in values to be clinically significant.However, it is unknown when these readings were obtained in relation to the reported medical event.
|
|
Event Description
|
A high readings issue was reported with use of the abbott diabetes care (adc) device.Customer received unspecified higher sensor scan results when compared to unspecified readings obtained on built-in meter.As a result, customer experienced "feeling of hunger", "shaking", and received third-party treatment of "juice and sugar" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.Built-in result(s) of 104 mg/dl and [comparison type] result(s) of 428 mg/dl were plotted on a parkes error grid and fell into the "d" zone, showing the difference in values to be clinically significant.However, it is unknown when these readings were obtained in relation to the reported medical event.
|
|
Manufacturer Narrative
|
Sensor (b)(6)has been returned and investigated.The sensor plug is fully seated and no issues were observed on sensor patch.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly.Observed deformed snaps upon visual inspection of the plug assembly.The current was applied to the sensor to perform accuracy testing while in the test fixture.All results were within specification.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.Therefore, issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high readings issue was reported with use of the abbott diabetes care (adc) device.Customer received unspecified higher sensor scan results when compared to unspecified readings obtained on built-in meter.As a result, customer experienced "feeling of hunger", "shaking", and received third-party treatment of "juice and sugar" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.Built-in result(s) of 104 mg/dl and [comparison type] result(s) of 428 mg/dl were plotted on a parkes error grid and fell into the "d" zone, showing the difference in values to be clinically significant.However, it is unknown when these readings were obtained in relation to the reported medical event.
|
|
Search Alerts/Recalls
|
|