• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITRON 300 SERIES G310SCLR; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITRON 300 SERIES G310SCLR; SCALER, ULTRASONIC Back to Search Results
Catalog Number G310SCLR
Device Problems Overheating of Device (1437); No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
While using a cavitron 300 series g310, they allege that water is not running through and the handpiece is getting hot, no injury resulted.
 
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
Edz dsp/if/ic/scan board shorted screen does not show power setting.Presetting button shows 00 on all of them.Debris buildup in the water filter.Replaced damaged/worn components and recalibrated unit to factory specs.Tech:water flow checked fine 5cc on min and 70cc on max which is good.No sterimate received for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON 300 SERIES G310SCLR
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key18118451
MDR Text Key327909182
Report Number2424472-2023-00053
Device Sequence Number1
Product Code ELC
UDI-Device IdentifierD00382703121
UDI-PublicD00382703121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberG310SCLR
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/10/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-