• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY C PHOSPHORUS REAGENT KIT; PHOSPHOMOLYBDATE (COLORIMETRIC), INORGANIC PHOSPHORUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY C PHOSPHORUS REAGENT KIT; PHOSPHOMOLYBDATE (COLORIMETRIC), INORGANIC PHOSPHORUS Back to Search Results
Catalog Number 08P40-30
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2023
Event Type  malfunction  
Event Description
The customer observed falsely depressed alinity c phosphorus results generated on the alinity c processing module.The following data was provided (reference range 2.3 - 4.7 mg/dl): (b)(6) initial result 8.2 mg/dl, repeated 10.2 mg/dl.No impact to patient management was reported.
 
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed falsely depressed alinity c phosphorus results generated on the alinity c processing module.The following data was provided (reference range 2.3 - 4.7 mg/dl): sid (b)(6) initial result 8.2 mg/dl, repeated 10.2 mg/dl.No impact to patient management was reported.
 
Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available; however, as part of troubleshooting, the customer repeated the same sample on another analyzer with the same lot of reagent.The rerun produced acceptable results.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.The ticket search by lot indicates that the complaint lot performs as expected for this product.Ticket trending review did not identify any trends.Device history review did not identify any non-conformances or deviations with the complaint lot.Labeling was reviewed and sufficiently addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the alinity c phosphorus reagent lot 82483un23 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY C PHOSPHORUS REAGENT KIT
Type of Device
PHOSPHOMOLYBDATE (COLORIMETRIC), INORGANIC PHOSPHORUS
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18121946
MDR Text Key328151912
Report Number3002809144-2023-00466
Device Sequence Number1
Product Code CEO
UDI-Device Identifier00380740135133
UDI-Public00380740135133
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/02/2024
Device Catalogue Number08P40-30
Device Lot Number82483UN23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY C PROCESSING MODU, 03R67-01, (B)(6).; ALNTY C PROCESSING MODU, 03R67-01, (B)(6).
-
-