• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON SYRINGE 10ML SALINE FILL CHINA SP; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON SYRINGE 10ML SALINE FILL CHINA SP; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306595
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4): initial mdr submission.A follow up mdr will be submitted if additional information becomes available.Patient problem code: f26 ¿ no health consequences or impact.Device problem code: a040101 ¿ fracture.
 
Event Description
2023.10.18 patient's postoperative catheter closure was performed using a prefilled catheter flosser, which was used normally.No other devices were used at the same time.During use, a crack was found at the needle stem and fluid was spilled, which was immediately replaced and used normally.
 
Event Description
(b)(4) follow up, on (b)(6) 2023 patient's postoperative catheter closure was performed using a prefilled catheter flosser, which was used normally.No other devices were used at the same time.During use, a crack was found at the needle stem and fluid was spilled, which was immediately replaced and used normally.
 
Manufacturer Narrative
(b)(4) follow up, a device history record review was completed by our quality engineer team for provided material number 306595 and lot number 2227182.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 10ML SALINE FILL CHINA SP
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
helen cox
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18125297
MDR Text Key329433903
Report Number1911916-2023-00841
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K003553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number306595
Device Lot Number2227182
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-