• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1711K
Device Problems Excess Flow or Over-Infusion (1311); Patient Device Interaction Problem (4001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2023
Event Type  malfunction  
Manufacturer Narrative
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Select patient information cannot be provided due to regional privacy regulations.The reported device is not marketed in the united states, but it is a same/similar device to one that is marketed outside the united states.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer reported that the customer was unable to turn on the extended bolus option.No further details were provided.Troubleshooting was not performed.No harm requiring medical intervention was reported.The customer will discontinue to use the insulin pump and will be returned for analysis.
 
Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Information has been corrected which was not correct in the initial report.The information has been provided in section h6 (medical device problem code) with this report.
 
Manufacturer Narrative
Pump was received with a critical pump error (open book image) alarm.Unable to perform the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement test, active current measurement test and self-test due to critical pump error (open book image) alarm.Pump was cut open to perform visual inspection and found moisture damage on the pcba 1, pcba 2 and force sensor.Successfully downloaded history files and traces using thus.Critical pump error (open book image) alarm was triggered by a pump error 35 fatal alarm confirmed in the history file and traces on 03/25/2023 20:48:00.000 due to moisture damage on the force sensor.Test p-cap and reservoir locked properly into reservoir compartment during testing.The following were noted during visual inspection: pillowing keypad overlay, stained keypad overlay and scratched case.Critical pump error (open book image) alarm confirmed due to pump error 35 alarm.Pump error 35 alarm confirmed due to moisture damage on the force sensor.Pump exposed to moisture confirmed at pcba 1, pcba 2 and force sensor.Cosmetic damages were noted during visual inspection.Unable to confirm extended bolus option due to critical pump error (open book image) alarm.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1711K
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
lusine boyadzhyan
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key18128026
MDR Text Key328037481
Report Number2032227-2023-301346
Device Sequence Number1
Product Code OYC
UDI-Device Identifier000000763000317089
UDI-Public(01)000000763000317089
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1711K
Device Catalogue NumberMMT-1711K
Device Lot NumberHG61VNG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-