MAUDE Adverse Event Report: VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number PRT-01185-002
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/09/2021
Event Type  malfunction  
Manufacturer Narrative
H6: the device was evaluated by ventec where the reported issue of it displaying a patient circuit disconnect alarm could not be confirmed.Proper device operation was confirmed through functional and performance testing.The cause of the reported issue could not be determined.
 
Event Description
It was reported that the device needed service.The device had previously been evaluated by an authorized third party service provider (asp).During the investigation, ventec observed that the asp had made a note in their service report stating that the device had provided an alarm due to patient circuit disconnect.This information had not been previously reported to ventec.There was no patient use associated with the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
2603 s goyer rd
kokomo IN 46902
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key18134355
MDR Text Key328130765
Report Number3013095415-2023-00718
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00850018761154
UDI-Public0100850018761154
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberPRT-01185-002
Device Catalogue NumberPRT-01185-002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2022
Date Manufacturer Received10/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-