Model Number HL-90
Device Problem Power Problem (3010)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the power cord terminals housing was defective.Patient involvement unknown.
 
Manufacturer Narrative
Other text: b3: date of event and d4: udi number is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Email is: (b)(6).Evaluation codes: updated.Device evaluation: one device was returned for investigation.Visual inspection found a broken pole clamp and cracked tank cover.Outdated printed circuit board (pcb) and power switch.Functional testing found the device has no power when plugged into power source and the power switch turned on; confirming the customer complaint.Root cause was attributed to the terminal strip on the pcb where the power cord wires are attached being damaged.What caused the damage could not be established.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key18136275
MDR Text Key328372615
Report Number3012307300-2023-10502
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-