LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/16/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that a heater-coolersystem 3t was contamination by mbc.There is no patient involvement.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no known patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in uk.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
H10: through follow-up communication under previous complaint from the same hospital, livanova learned that the device was cleaned regularly as per the instructions for use, however when the device is not used it is stored full of water and h2o2 is not checked daily.This is not in accordance with device instructions for use and this deviation may have led/contributed to the reported contamination.
|
|
Event Description
|
See initial report.
|
|