SPATZ FGIA INC. SPATZ3 ADJUSTABLE BALLOON SYSTEM; GASTRIC BALLOON Back to Search Results
Model Number A-SP3-03K
Device Problem Deflation Problem (1149)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2023
Event Type  malfunction  
Event Description
The doctor wanted to fill the balloon with liquid, but only a small part of liquid went into the balloon, the bigger part went into the patient's stomach.
 
Manufacturer Narrative
To date, spatz fgia inc.Has not received the product for evaluation, therefore no analysis or testing has been done.A review of the device labeling notes the following: each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed to report to physicians immediately regarding any and all change of symptoms.Symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur should be reviewed with patient, and patients should be advised to contact his/her physician immediately upon the onset of such symptoms.Patients reporting loss of satiety, increased hunger and/or weight gain should be examined radiographically and/or endoscopically, as this is indicative of a balloon deflation.It is necessary to replace a balloon which has spontaneously deflated.Complications- possible complications of the use of the spatz3 adjustable balloon system include: balloon deflation and subsequent replacement.
 
Manufacturer Narrative
The device was returned to spatz fgia inc.For analysis on 11/26/2023 and an evaluation was completed.The testing revealed a linear cut at the inflation tube area that probably was caused during the implantation procedure.A review of the device labeling notes the following: each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed to report to physicians immediately regarding any and all change of symptoms.Symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur should be reviewed with patient, and patients should be advised to contact his/her physician immediately upon the onset of such symptoms.Patients reporting loss of satiety, increased hunger and/or weight gain should be examined radiographically and/or endoscopically, as this is indicative of a balloon deflation.It is necessary to replace a balloon which has spontaneously deflated.Complications- possible complications of the use of the spatz3 adjustable balloon system include: balloon deflation and subsequent replacement.
 
Event Description
The doctor wanted to fill the balloon with liquid, but only a small part of liquid went into the balloon, the bigger part went into the patient's stomach.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPATZ3 ADJUSTABLE BALLOON SYSTEM
Type of Device
GASTRIC BALLOON
Manufacturer (Section D)
SPATZ FGIA INC.
1801 s perimeter rd, ste 130
fort lauderdale FL 33309
Manufacturer (Section G)
SPATZ FGIA INC.
1801 s perimeter rd, ste 130
fort lauderdale FL 33309
Manufacturer Contact
pnina polishuk
1801 s perimeter rd, ste 130
fort lauderdale, FL 33309
MDR Report Key18138497
MDR Text Key328241545
Report Number3012638928-2023-02931
Device Sequence Number1
Product Code LTI
UDI-Device Identifier860005178810
UDI-Public(01)860005178810(17)2024-08-10(10)220810
Combination Product (y/n)N
Reporter Country CodeHU
PMA/PMN Number
P190012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA-SP3-03K
Device Lot Number220810
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2023
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer Received10/19/2023
Supplement Dates FDA Received12/12/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-