According to the log file the ventilator board repeatedly stopped its can/sa-bus communication, which interrupted the communication between user interface (mobi) and ventilator as well as the one to the gas mixer, as this communication including settings and status feedback is routed via the ventilator.Resulting from the loss of communication, ventilator and gas mixer initiated an emergency shutdown to monitoring mode while the mobi alarmed for gas + vent fail including the prompts mentioned above.It can be concluded that the reported event was caused by an internal communication problem of the ventilator¿s cpu board.Manual ventilation, using the device breathing bag while setting an adequate flow via the o2 flow control valve (incl.Agent) remained possible.Based on the fact, that the same pcb is also assembled in the gas mixer without showing this symptom, a general design failure can be excluded.It cannot be excluded that the reported phenomenon was triggered by electromagnetic radiation or electrostatic discharge of the user during interactions with the device.Apollo is designed, tested and certified to withstand emc/esd influence in conformance to the relevant standard (iec 60601-1-2).The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|