Model Number N/A |
Device Problems
Unstable (1667); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported during preventative maintenance, the device had an unstable motor speed and worn screws.There was no patient involvement.There are no adverse events associated with this malfunction.Due diligence is complete, no further information is available.
|
|
Manufacturer Narrative
|
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.G2: foreign: poland.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the screws were worn and the motor rpms were unstable.The motor and screws were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional information available.
|
|