Model Number 11500AJ23
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Endocarditis (1834); Insufficient Information (4580)
Event Date 06/02/2023
Event Type  Injury  
Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
Edwards received information through its implant patient registry that a 23mm 11500aj aortic valve was explanted after a duration of one (1) months due to unknown reason.Another same model 23mm 11500aj aortic valve was implanted in replacement.
 
Event Description
Edwards received information through its implant patient registry that a 23mm 11500aj aortic valve was explanted after a duration of forty-two (42) days due to prosthetic valve endocarditis.Another same model 23mm 11500aj aortic valve was implanted in replacement.The device was not returned for evaluation due to infection.Type and source of infection were not reported.There was no allegation of device malfunction.
 
Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Prosthetic endocarditis, with or without vegetation, of valves and annuloplasty rings is a serious complication of cardiac valve replacement and valve repair surgeries despite improvements in prostheses types, surgical techniques, and infection control measures.Prosthetic endocarditis occurring within 60 days of valve or ring implantation generally reflects contamination arising in the perioperative period.There are many opportunities for organisms to seed a prosthesis perioperatively, most of which probably occurs intraoperatively.Besides the patient's own skin and access lines, several other important modes of contamination have been recognized including air in the operating room, the coronary suction devices used during surgery and the heart-lung bypass machine, faulty technique during cardiac output measurements, and the prosthesis itself.In early cases of prosthetic endocarditis, subsequent infections are almost universally related to contamination at the time of surgery.If there were ever non-conformances in the sterility or packaging processes, they would most likely manifest in the early post-operative period.Since there are multiple modes of contamination other than the prosthesis itself, and no indication or allegation that the patient's infection was device related, the event was likely due to patient and/or procedural-related factors.An engineering evaluation is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and there is no evidence of a product failure with regard to design, reliability, or use error.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A capa/scar/pra is not required as there are no confirmed product or labeling nonconformances and no other triggers are met.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRIS RESILIA AORTIC VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key18141576
MDR Text Key328185285
Report Number2015691-2023-17507
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number11500AJ23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/30/2023
Date Device Manufactured11/25/2022
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Life Threatening;
Patient Age82 YR
Patient SexMale
-
-