H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Prosthetic endocarditis, with or without vegetation, of valves and annuloplasty rings is a serious complication of cardiac valve replacement and valve repair surgeries despite improvements in prostheses types, surgical techniques, and infection control measures.Prosthetic endocarditis occurring within 60 days of valve or ring implantation generally reflects contamination arising in the perioperative period.There are many opportunities for organisms to seed a prosthesis perioperatively, most of which probably occurs intraoperatively.Besides the patient's own skin and access lines, several other important modes of contamination have been recognized including air in the operating room, the coronary suction devices used during surgery and the heart-lung bypass machine, faulty technique during cardiac output measurements, and the prosthesis itself.In early cases of prosthetic endocarditis, subsequent infections are almost universally related to contamination at the time of surgery.If there were ever non-conformances in the sterility or packaging processes, they would most likely manifest in the early post-operative period.Since there are multiple modes of contamination other than the prosthesis itself, and no indication or allegation that the patient's infection was device related, the event was likely due to patient and/or procedural-related factors.An engineering evaluation is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and there is no evidence of a product failure with regard to design, reliability, or use error.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A capa/scar/pra is not required as there are no confirmed product or labeling nonconformances and no other triggers are met.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.
|