DEPUY IRELAND - 3015516266 SMARTSET GHV GENTAMICIN 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3095040
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cyst(s) (1800)
Event Date 09/23/2013
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).No 510k as device is not marketed in the united states under this product code, but the same/similar product is marketed in the us under a different product code.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Subject id: (b)(6).Study no: (b)(6).Clinical adverse event received for baker's cyst.Event is not serious and is considered moderate.Event is possibly related to procedure.Event is not related to device.Date of implant: (b)(6) 2013.Date of event: (b)(6) 2013.(left knee).Treatment: us guided aspiration and steroid injection.
 
Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A complaint database search did find additional results but no related reports against the provided product code and lot number combination(s).Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GHV GENTAMICIN 40G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool
UK  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18147061
MDR Text Key328225203
Report Number1818910-2023-23296
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
OUS - SIMILA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2015
Device Catalogue Number3095040
Device Lot Number3572269
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATTUNE CR FEM LT SZ 8 CEM.; ATTUNE CR RP INSRT SZ 8 8MM.; ATTUNE MEDIAL DOME PAT 41MM.; ATTUNE RP TIB BASE SZ 7 CEM.
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight74 KG
-
-