• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXALT MODEL D SINGLE-USE DUODENOSCOPE; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXALT MODEL D SINGLE-USE DUODENOSCOPE; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00542421
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Manufacturer Narrative
Block h6: problem code a06 captures the reportable event of loss of visualization inside the patient.
 
Event Description
It was reported to boston scientific corporation that an exalt model d single-use duodenoscope was used on an endoscopic retrograde cholangiopancreatography (ercp) used to treat stones performed on october 24, 2023.During cannulation, the image was flickering.At one point, the screen went red and then visualization was completely lost.The scope was unplugged and plugged back in but the issue persisted.The procedure was successfully completed using a reusable scope.There were no patient complications as a result of this event.Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report number 3005099803-2023-06236 for the exalt model d controller.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXALT MODEL D SINGLE-USE DUODENOSCOPE
Type of Device
DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18147678
MDR Text Key328261727
Report Number3005099803-2023-06221
Device Sequence Number1
Product Code FDT
UDI-Device Identifier08714729993605
UDI-Public08714729993605
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00542421
Device Catalogue Number42421
Device Lot Number0032465845
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-