Model Number 16-02-85 |
Device Problem
Insufficient Cooling (1130)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/21/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t was not cooling down on the patien side more then 25 degrees post procedure.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in usa.A livanova field service engineer reported also the customer complained of getting a press switch tripped (e27) and display button where not responding and needed to be replaced.The display keyboard, the processor board and patient circuit bridge were replaced.All components have been inspected and tested according unit returned to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: complaints database analysis revealed that no similar event on this device occurred since its installation in 2020.Based on the collected information, the root cause of the reported event of insufficient cooling can be traced back to a defective patient bridge.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event, however there is no concerning trend for this kind of failure.
|
|