• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS Back to Search Results
Model Number A42011A
Device Problems Break (1069); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
The customer reported that his olympus resection sheath had button damage and a loose ceramic tip.According to the initial reporter, this event had no patient/procedure involvement.
 
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.The ceramic tip was found damaged.Additionally, the black cap was missing and both the spring housing as well as the sealing ring were damaged.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS
Type of Device
SHEATH
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18148334
MDR Text Key328278683
Report Number9610773-2023-03315
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761023658
UDI-Public04042761023658
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA42011A
Device Lot Number108W-0039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2023
Was the Report Sent to FDA? No
Date Manufacturer Received11/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-