DEPUY ORTHOPAEDICS INC US RETAINING STEM INSERTER; HIP INSTRUMENTS : INSERTION DEVICES Back to Search Results
Catalog Number 259807570
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that the inserter/extractor is cross-threaded and needs to be replaced.No time was added to surgery.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that the inserter/extractor is cross-threaded and needs to be replaced.No time was added to surgery.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the proximal threaded area of the shaft was heavily damaged.This damage can be attributed to a combination of heavy usage and impacting the threaded rod weldment before is fully threaded in to the retaining stem inserter body.The overall complaint was confirmed as the observed condition of the retaining stem inserter would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicates that there was no harm to the patient.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETAINING STEM INSERTER
Type of Device
HIP INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18148343
MDR Text Key328278710
Report Number1818910-2023-23423
Device Sequence Number1
Product Code HWR
UDI-Device Identifier10603295148548
UDI-Public10603295148548
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number259807570
Device Lot NumberAB5383794
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-