ABBOTT ENDURITY¿ CORE DR, JPD; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2152
Device Problem No Pacing (3268)
Patient Problem Arrhythmia (1721)
Event Date 11/01/2023
Event Type  Injury  
Event Description
A loss of pacing was observed on the device resulting in cardiac decompensation.The event was resolved by explanting and replacing the device.The patient was in stable condition following the procedure.
 
Manufacturer Narrative
The reported event of no pacing output was confirmed.The device was received with no output and no telemetry communication.Visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery voltage was found normal.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the current drain was found normal.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on(b)(6) 2022 for a subset of devices distributed and implanted outside of the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY¿ CORE DR, JPD
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18151595
MDR Text Key328310559
Report Number2017865-2023-52685
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberPM2152
Device Lot NumberP000097509
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-