INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1114
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that after repair of the mayfield infinity skull clamp (a1114), the customer reused the mayfield for a surgery.Before the surgery, the two arms did not slide, not allowing the patient's skull to be fixed for the surgery.The clinician took a video of the malfunction which was sent to our service and repair site.The surgeon continued the surgery after 90 minutes delay due to the switching to another mayfield they had in stock.No injury to the patient was observed.The exact event date is unknown.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
N/a.
 
Manufacturer Narrative
The mayfield infinity skull clamp (a1114) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device found no device deficiencies that would have contributed to the reported complaint.The video sent to us by the customer was very helpful, but we could not replicate the issue while testing the unit according to the manufacturer's specifications.No maintenance is required at this time.However, the unit subjected to a successful functional check and will be marked with an instance number.Root cause - the complaint is not confirmed.The returned unit is in good working order and meets all manufacturer¿s specifications, and no maintenance is required.Probable root cause is improper use.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18152218
MDR Text Key329576177
Report Number3004608878-2023-00205
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K051440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1114
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-