The event occurred at 5:30am involving a 14 cm (5.5") smallbore ext set w/microclave® clear, clamp, rotating luer where the customer reported a presence of blood around a patient during gravity infusion.A pediatric nurse observed the absence of tubing at the catheter placed on the child's forearm.The catheter's extension broke or detached at a weld joint, causing the blood to flow onto the floor, it appeared that the valve detached from the catheter.The ongoing intravenous solution was isopedia at a flow rate of 60 ml/h.The medical device was connected to a pink 20g catheter.The catheter was inserted from tuesday night, (b)(6) 2023 to wednesday, (b)(6) 2023, at 23:00, adding to it the extension 011-mc33793, a valve with extension and a standard tubing for the volumat pump during gravity infusion.At 05:00 am the antibiotics were connected.At 05:38 am the patient rang the bell to report blood on the bed.The nurse immediately redressed the bandage and placed a new catheter extension.During handling of the defective device, the nurse used biseptine.There were no changes in the patient's condition before, during or after the incident.No one was harmed when the incident occurred, no adverse event have been noted, there was, no delay in therapy, the therapy was successful after the device has been changed and the patient has received the medication planned dose.There was a blood loss, but there was no quantifiable clinically significant blood loss noted.Regarding the volumat infusion pump, the customer stated no alarm has been showed/ringed by the pump.The 20g pink catheter has been connected correctly as per manufacture's guide.There was patient involvement but no adverse events/human harm.
|
D9 - date returned to mfg on 12/19/2023.The reported complaint of breakage was confirmed on the returned set.During visual inspection, the pressure tubing was received broken/cut off from the microclave.When the returned set was microscopically examined, a part of the tubing was found inside the microclave luer.The probable cause of the tubing breakage had occurred due to unintentional twisting or pulling force applied on the tubing during use.The device history report (dhr) for lot # review could not be conducted because no lot number(s) was/were identified.
|