Visual examination revealed no visual anomalies observed.3-way stop cock was returned attached to device.The polarsheath was functionally tested in attempt to recreate the visible air allegation seen in the field.The functional tests include an aspiration, hemostasis, and positive air pressure test.These tests evaluate the performance of the hemostatic valve, some of which are under conditions more rigorous than a clinical setting.The sheath failed all standard manufacturing testing, with a leak in the pressure decay test, air in the flush line during the aspiration test and dropping pressure while pressurized at 5.5 psi in hemostasis testing.The sheath failed standard manufacturing testing when air was observed in the flush line during the aspiration test.The sheath handle was disassembled to find the area where the leak occurred.Analysis found a leak coming from the hemostatic valve at the proximal end.There was a tear in the hemostatic valve due to an off-center puncture which resulted in the polarsheath leaking.The ifu 51331210 states "do not aspirate via the side port if the sheath lumen is occupied (i.E., by the dilator or components of the cryoablation catheter) as the aspiration may draw air across the sheath valve into the polarsheath." per the complaint summary, "when the balloon catheter was inserted after sheath insertion, a reverse blood flow confirmation was performed." this confirms an attempt to aspirate the polarsheath with the catheter occupying the sheath lumen.
|