VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number PRT-01100-000
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2023
Event Type  malfunction  
Manufacturer Narrative
H6: the device was evaluated by ventec where the reported issue of displaying a patient circuit disconnect alarm was confirmed.Ventec replaced the internal flow transducer (ift) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the ift.
 
Event Description
It was reported that the device needed service.During the device's evaluation, ventec observed that it was displaying a patient circuit disconnect alarm.There were no reports of patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key18156759
MDR Text Key328991712
Report Number3013095415-2023-00744
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00855573007747
UDI-Public0100855573007747
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPRT-01100-000
Device Catalogue NumberPRT-01100-000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2023
Date Manufacturer Received10/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-