Model Number CV-190
Device Problems No Display/Image (1183); Moisture Damage (1405)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.There is a loose scope socket mechanical lock on the video connector socket that does not secure any paddles (or camera heads).All the paddles were loose and disconnected, when slightly moved caused noise or loss of image.The additional findings were as follows: that the power switch needed to be replaced due to a substance spilling on it.The unit was cracked and damaged on the front panel, the top cover needs to be replaced due to poor appearance, and a software 3.01.00 upgrade was needed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the video system center had no image.The issue was found at maintenance.There were no reports of patient involvement or patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.It has been over 7 years, since the subject device was manufactured.A definitive root cause could not be determined, for the events reported.Based on the results of the investigation, about the "no image".It is likely, the following led to the malfunction loosening of the scope socket.And for "the material spilled" event, there is not any probable cause.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18156880
MDR Text Key329777477
Report Number3002808148-2023-13015
Device Sequence Number1
Product Code FET
UDI-Device Identifier04953170298622
UDI-Public04953170298622
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-