• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Device Problem Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/18/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: mustang 10.0 x 60, 75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A balloon longitudinal tear was identified beginning approximately 25mm distal of the proximal markerband and extending approximately 11mm distally across the balloon material.The rated burst pressure for this device is 14 atmospheres.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination found no damage to the tip of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged in the correct position on the device.
 
Event Description
Reportable based on device analysis completed on 15-nov-2023.It was reported that advancing difficulties were encountered.Vascular access was obtained via contralateral approach.The 100% stenosed target lesion was located in the severely tortuous common iliac artery.A 10.0 x 60, 75cm mustang balloon catheter was advanced for dilation.However, during the procedure, the device did not move in the opposite direction during the contralateral approach process.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed balloon longitudinal tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18162328
MDR Text Key328511351
Report Number2124215-2023-64770
Device Sequence Number1
Product Code FGE
UDI-Device Identifier08714729794004
UDI-Public08714729794004
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K141521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0030382053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-