ETHICON ENDO-SURGERY, LLC. FLEX 60 ARTICULATING; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number EC60A
Device Problems Mechanical Problem (1384); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported that during an unk procedure, the device opening button was stuck and not working.No patient consequences were reported.
 
Manufacturer Narrative
(b)(4).Date sent: 11/17/2023.D4: batch # 639a19.B3: exact event date unk, entered 1/1/2023 as only the year was provided.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one ec60a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties noted.The staple line and cut line were complete and the staples meet the staple release criteria.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: if the jaws do not automatically open after the anvil release button is pressed, first ensure that the knife is in the retracted position by verifying that the stroke count indicator displays ¿0¿ and knife direction indicator points towards the proximal side of the instrument, or that the knife blade indicator is in the home position.If the stroke count indicator or knife blade indicator is not in the home position, push the red manual knife reverse switch downward to reverse the knife motion and squeeze the firing trigger, completely until it rests on the closing trigger.Press the anvil release button.If the jaws do not open at this point, then gently pull the closing trigger, upward (away from the handle) until both firing and closing triggers return to their original positions.If the instrument locks out, the stroke count indicator will display a lockout symbol.Stop and push the red manual knife reverse switch downward to reverse the knife motion; the knife direction indicator will display an arrow pointing towards the proximal end of the instrument to indicate the knife is in the return mode.Squeeze firing trigger (2) completely until it rests on the closing trigger.The stroke count indicator will display ¿0¿ to indicate the knife has returned to its home position.Then press the anvil release button to release the jaws from the tissue.Close the instrument jaws by pressing the closing trigger (1), remove the instrument, and replace the cartridge.Firing through the lockout mechanism will break the instrument.The event described could not be confirmed as the device performed without any difficulties noted and no cartridge was returned for analysis.Device history review a manufacturing record evaluation was performed for the finished device batch number 639a19, and no non-conformances were identified.Additional information was requested and the following was obtained: was the device locked on tissue with the jaws closed? if yes, how was the device removed? what troubleshooting steps were attempted to open the device (e.G.Squeezing the closing trigger while simultaneously depressing the anvil release button, knife reverse switch, manual override)? did the jaws eventually open?.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEX 60 ARTICULATING
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18164195
MDR Text Key328442289
Report Number3005075853-2023-08557
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036001706
UDI-Public10705036001706
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K081146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEC60A
Device Lot NumberX94A9E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2023
Date Manufacturer Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-