The customer took samples of dust in the laboratory and also swabbed the area around the instrument, as well as sample/reagent probe.The samples were tested at a nearby facility and each one resulted as greater than the measuring range for testosterone.The field service engineer determined the cause of the event was the testosterone contamination, likely coming from the hyase product used in the ivf lab on site.He replaced the sample/reagent probe and pressure sensor tubing.He performed sample/reagent probe, sipper, and gripper adjustments.He replaced the measuring cell and syringe assembly and ran performance testing that passed.As the qc recovery was within the acceptable range, there was no indication of a performance issue with the reagent.
|