• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL COSTA RICA LTD. UNSPECIFIED PLUM TUBING; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL COSTA RICA LTD. UNSPECIFIED PLUM TUBING; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned for evaluation; however, testing has not yet been completed.
 
Event Description
It was reported that, on an unknown date, a plum intravenous (iv) tubing set fell apart, during patient use, where it connects to the cartridge.There was no patient harm reported.
 
Manufacturer Narrative
Received one (1) used list #unknown plum set for inspection.The tubing was separated from the inlet port of the cassette.Examination of the bond showed the tubing remained bonded inside of the port and the tubing was torn apart.Evaluation under uv light showed sufficient solvent.The reported complaint can be confirmed.The probable cause is due to an unintentional pulling force during use.A lot review could not be conducted because no lot number(s) was/were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED PLUM TUBING
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18166919
MDR Text Key329094297
Report Number9615050-2023-00658
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00(01)(17)(10)
UDI-Public(01)(17)(10)
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received12/11/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-