ICU MEDICAL COSTA RICA LTD. UNSPECIFIED PLUM TUBING; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number UNKNOWN |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned for evaluation; however, testing has not yet been completed.
|
|
Event Description
|
It was reported that, on an unknown date, a plum intravenous (iv) tubing set fell apart, during patient use, where it connects to the cartridge.There was no patient harm reported.
|
|
Manufacturer Narrative
|
Received one (1) used list #unknown plum set for inspection.The tubing was separated from the inlet port of the cassette.Examination of the bond showed the tubing remained bonded inside of the port and the tubing was torn apart.Evaluation under uv light showed sufficient solvent.The reported complaint can be confirmed.The probable cause is due to an unintentional pulling force during use.A lot review could not be conducted because no lot number(s) was/were identified.
|
|