Catalog Number 71118200
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2023
Event Type  malfunction  
Manufacturer Narrative
Internal complaint number: (b)(4).
 
Event Description
It was reported that during a tibial pseudoarthrosis procedure, at the end of the last reaming of the spinal canal, when removing the flexible shaft w/cir connector, it had broken.The procedure was resumed, without any delay, using the same device.There was no breakage inside the patient and no injury was reported as a consequence of this issue.
 
Manufacturer Narrative
The device was not returned for evaluation; therefore, a device analysis could not be performed.However, the photographs were reviewed and revealed that only the tip and part of the shaft are shown and there are no signs of breakage.As the whole device is not shown, it can not be confirmed if it is broken or not.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18166920
MDR Text Key328514080
Report Number1020279-2023-02260
Device Sequence Number1
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public03596010436528
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K121714
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number71118200
Device Lot Number12GM09901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
-
-