Model Number N/A
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Event Description
It was reported that there was an instrument that was worn.There was no patient involvement.Attempts have been made and there is no further information at this time.
 
Manufacturer Narrative
(b)(4).H3: product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).Proposed annex g code: mechanical (g04) - drill reported event was confirmed as visual examination of the returned product/provided pictures identified the device was returned showing signs of repeated use.The baseplate reamer has nicks and gouges on the tip of the reamer as well as galling on the shaft of the reamer.The ao connector end has some scratching and gouges.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPR RVS BSPLT RMR 25MM SHORT
Type of Device
EXTREMITIES INSTRUMENTS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18169907
MDR Text Key328535728
Report Number0001825034-2023-02718
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00887868470110
UDI-Public(01)00887868470110(10)196250
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110029136
Device Lot Number196250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Was the Report Sent to FDA? No
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-