Catalog Number RTLR180343 |
Device Problem
Sparking (2595)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
It was reported that the patient¿s liberty select cycler sparked during preparation of their peritoneal dialysis (pd) treatment.The cycler sparked from the back of the cycler when the cycler was plugged in at home.The patient noted that they just came back from a trip where the cycler was carried in a medical bag.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.Additional information requested but has not been obtained.
|
|
Event Description
|
It was reported that the patient¿s liberty select cycler sparked during preparation of their peritoneal dialysis (pd) treatment.The cycler sparked from the back of the cycler when the cycler was plugged in at home.The patient noted that they just came back from a trip where the cycler was carried in a medical bag.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was available.Additional information requested but has not been obtained.
|
|
Manufacturer Narrative
|
Additional information provided in d9 and h3.Plant investigation: a visual inspection of the returned cycler exterior showed signs of physical damage on power entry module.An internal inspection of the cycler found quick connect crimps disconnected from power terminals and in contact with each other causing cycler to short.Pitting was observed on the quick connect crimps where the short was occurring.A known good power entry module was temporarily installed.Plug unit in, power up check passed.No audible alarms or alarm screen were displayed.Voltage test passed.Valve actuation test passed.System air leak test passed.Post- accelerated stress test 2hours 15mins 8500ml simulated treatment was performed without any failures or problems.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed.The cycler was refurbished following the evaluation.
|
|
Search Alerts/Recalls
|